Pre procedure notes | Familiarise yourself with the equipment and manufacturer instructions on calibration or calibration verification. Spirometer equipment ideally is ISO 26782 compliant and calibration syringe has an in-date certificate and has been pre checked for leaks (daily inspection and monthly leak check must be documented). If using software consult the manufacturer instructions for use. Checks all parts are in good working order before this procedure. Note that room temperature and syringe temperature may negatively affect results. Refer to latest national and international standards. |
Calibration types | A calibration verification is the procedure used to validate that the device is within calibration limits (i.e., =/-3% [accuracy tolerance, =/-2.5% for spirometers plus =/-0.5% for calibration syringes]). In calibration mode the equipment will exclude the BTPS factor to the results. |
Note on pre-calibrated sensors | Precalibrated sensors must undergo calibration verification. Manufacturers must specify the action to be taken if a pre-calibrated device fails the calibration verification. |
Setup procedure | Connect the calibrator (3 L Syringe) securely to the spirometer sensor without causing allowing air to flow through the device. If variable flow is detected during the zero-flow setting procedure or if the zero level has changed significantly, the zero-flow setting procedure must be repeated. |
Perform procedure | Fill the syringe completely with air and attach the calibration syringe firmly to the equipment sensor (always use a filter during the calibration of flow sensing equipment). |
Empty the syringe smoothly and completely until a soft click is heard, continuously observe the syringe strokes on the equipment screen. Do not bang the syringe when emptying it, to avoid any damage. Tip: follow the screen or software instructions to aid performance of this procedure. Linearity of the system is assessed as follows: The calibrating manoeuver is performed three times at different flows. The recorded volume should be repeatable: when using a 3L syringe the largest - smallest measured volume should not deviate by more than 90 mL. Repeat above 3 to 5 times, you will be instructed to alter the speed of the air injections (slow, medium and fast) and must comply with the "multi flow" calibration standard (a range varying between 0.5 and 12 L/s with 3-L injection times between 0.5 and 6 s). The resulting simulated FVC values must fall within the 3% range. Each time stamped successful calibration must be stored either within the software or as a print-out. | |
Calibration result | The measured volume displacement should be within 2.91 and 3.09 at ATPS conditions. Results must be printed or stored for recall later. |
Documentation | A log of all quality control findings, repairs and adjustments, and hardware and software updates d Verification of reference value calculations after software updates. |
Spirometer | Device Quality Assurance Attention to equipment quality assurance and calibration is an important part of good laboratory practice The minimum requirements are as follows: maintenance of a log of calibration results documentation of repairs or other alterations that return the equipment to acceptable operation recording of dates of computer software and hardware updates or changes recording the dates equipment is changed or relocated (e.g., industrial surveys). Calibration verifications and quality control procedures must be repeated after any such changes before further testing begins. |
Table 1. Daily calibration procedure
Video 1. Spirometer calibration
Table 2. Spirometry test hygiene, infection protection and control steps (should comply with local policy)
Video 2. Hand washing
Video 3. Special considerations - CUT OUT THE PART ON CLEANING